Function |
Equipment |
Equipment Description |
Quantity Used |
Alarm Lamp System |
Alarm lamp |
Strip alarm lamp |
One set |
Central Power Distribution System |
Multifunctional power supply control system |
It supports two power supply modes, external power supply (to be connected with utility
power) and vehicle-mounted intelligent inverter power supply system, capable of automatic
switching between 12 V DC and 220 V AC power supplies. |
One set |
Medicine Cabinet System |
Medicine cabinet |
Light-weight environmental protection material |
One set |
Oxygen cylinder cabinet |
Light-weight environmental protection material |
One set |
Combined cabinet |
Light-weight environmental protection material |
One set |
Medical Floor |
Medical floor leather |
Resistant to acid, alkali, fire, skidding, static electricity |
One set |
Specialized Medical Equipment |
Oxygen cylinder |
10L |
Two sets |
Sputum aspirator |
|
One set |
Sphygmomanometer |
|
One set |
Ultraviolet disinfection lamp |
|
One set |
Oxygen supply system |
|
One set |
Stretcher System |
Deluxe automatic boarding stretcher + automatic boarding platform (including shovel
stretcher) |
Patient transferring |
One set |
Medical Cabin Lighting System |
Working lamp |
|
One set |
Stretcher-type Negative Pressure Isolation Cabin |
Stretcher-type Negative Pressure Isolation Cabin |
Power supply: rated voltage of AC 220 V, 50 HZ, internal power supply DC 12 V, input
power of 50 VA. Filtration power: filtration efficiency ≥99.99% for 0.3 μm particulate
aerosol. Negative pressure: A negative pressure of not less than 15 Pa shall be established
within 2 min when there is passenger in the ambulance. Ventilation rate: ≥75 L/min.
Effective power supply duration: lithium battery power supply, ≥2 h. Noise: ≤72 dB(A). |
One set |
Negative Pressure System |
Fan system |
Exhaust air rate: 200 m3/h-600 m3/h, 220 V, 300 W to ensure high
ventilation efficiency in the medical cabin, and to minimize the probability of cross
infection among medical staff during the process of treating and transferring patients with
infectious diseases. |
One set |
Filtration system |
Filtration power: filtration efficiency ≥99.99% for 0.3 μm particulate aerosol (the
national standard requiring ≥99.97%). The air in the van will be discharged after bio-safety
disposal. |
One set |